| Model Number |
270333 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while testing with the bd macro-vue¿ rpr card antigen suspension, discrepant results were obtained.The result was positive while confirmatory testing with a competitor's reagent produced a negative result.Positive result was reported out; there was no report of patient impact.The following information was provided by the initial reporter: customer in (b)(6) received a (1 : 32) titer for one of the patient samples while (b)(6) did the rpr test for this patient using competitor's reagents, and obtained a titer of (1 : 1); patient result was reported as (1 : 32);.
|
| |
|
Additional Manufacturer Narrative
|
|
Exact date of event is unknown.Information provided: "this test was done in (b)(6) 2021" a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary satisfactory results when retention samples were tested for performance using three (3) sera panel numbers of each reactivity.Control lot (1335592) customer batch number (1123493) serum no.11: all lots showed reactive results at 1:1 to 1:4 and reactive minimal at 1:8.Serum no; 16: all lots showed reactive results at 1:1 and reactive minimal at 1:2.Serum no; 17: all lots showed reactive results at 1:1 to 1:2 and reactive minimal at 1:4.Serum no¿s.29, 32, 34 all lots showed reactive minimal results at 1:1 to 1:2.Serum no¿s.3,7,10: all lots showed negative results.Complaint is unconfirmed based on retention samples and batch history record review.No returned goods received.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process cgg.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that while testing with the bd macro-vue¿ rpr card antigen suspension, discrepant results were obtained.The result was positive while confirmatory testing with a competitor's reagent produced a negative result.Positive result was reported out; there was no report of patient impact.The following information was provided by the initial reporter: customer in clinical labs of hawaii received a (1 : 32) titer for one of the patient samples while doh of hawaii did the rpr test for this patient using competitor's reagents, and obtained a titer of (1 : 1); patient result was reported as (1 : 32);.
|
| |
|