• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION POLARIS ULTRA; STENT, URETERAL Back to Search Results
Model Number 192-142
Medical Device Problem Code Detachment of Device or Device Component (2907)
Date of Event 09/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The returned polaris ultra ureteral stent was analyzed, and a visual evaluation noted that the suture hole torn and the shaft detached.The suture did not returned only the positioner returned.No other issues with the device were noted.The reported event was not confirmed.According to the product analysis, it is possible that operational factors, interaction of the device with the suture string such as entanglement, manipulation or excess of force applied during it use, could have caused the detached observed.Consequently, affect the performance of the device.In the other hand the report complaint of coil detachment of device or device component cannot be confirmed due to the coil was not detached, for this reason the report issue cannot be confirmed.Therefore, adverse event related to procedure is selected as the most probable root cause for the complaint.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported that a polaris ultra ureteral stent was used during a retrograde intrarenal surgery procedure, performed on (b)(6) 2021.During the procedure and inside the patient, when the physician pulled back the suture for stent deployment, it was noticed that the coil of the stent was broken.Another polaris ultra ureteral stent was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation finding of stent shaft broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARIS ULTRA
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13766426
Report Number3005099803-2022-01183
Device Sequence Number14460303
Product Code FAD
UDI-Device Identifier08714729124542
UDI-Public08714729124542
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K010002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2024
Device Model Number192-142
Device Catalogue Number192-142
Device Lot Number0026803872
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Initial Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
-
-