| Model Number |
PM2134 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
02/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2022-05427, related manufacturer reference number: 2017865-2022-05428.It was reported that the patient presented to the hospital for a check up with signs of infection.Upon examination of the system, it was noted that there was infection which was developed at the pocket incision site.The pacemaker, the right atrial lead and right ventricular lead were explanted on (b)(6) 2022.The patient was in stable condition throughout.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the product was removed due to infection.During analysis, the interrogation results were normal.Visual screen values were acceptable.Device image download was successful.Preliminary test results were acceptable.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The cause of infection could not be determined.
|
| |
|