• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941838350
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified mid right coronary artery.After a wire crossed the lesion and pre-dilatation was performed with non compliant balloons, a 3.50 x 38mm synergy drug-eluting stent was selected for use.When the physician removed the stent from the hoop, the stent delivery catheter was kinked.However, the technologists loaded the stent into the wire and as the physician inserted the device into the guide catheter, the shaft fractured about 100cm from the hub.The device was removed and the procedure was completed with another of the same device.No patient complications were reported.
 
Event or Problem Description
It was reported that shaft break occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified mid right coronary artery.After a wire crossed the lesion and pre-dilatation was performed with non compliant balloons, a 3.50 x 38mm synergy drug-eluting stent was selected for use.When the physician removed the stent from the hoop, the stent delivery catheter was kinked.However, the technologists loaded the stent into the wire and as the physician inserted the device into the guide catheter, the shaft fractured about 100cm from the hub.The device was removed and the procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Synergy xd mr us 3.50 x 38mm stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of tip damage.A visual and tactile examination of the hypotube identified a break on the hypotube shaft located at 67.6 cm distal to the distal end of strain relief as well as multiple kinks.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and tactile examination.No issues were noted.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13767256
Report Number2134265-2022-02919
Device Sequence Number5008142
Product Code NIQ
UDI-Device Identifier08714729981091
UDI-Public08714729981091
Combination Product (Y/N)Y
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/05/2023
Device Model NumberH7493941838350
Device Lot Number0027866989
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date05/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1