STRYKER ORTHOPAEDICS-MAHWAH HMRS AXIS PIN; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6366-9-220
Medical Device Problem Codes Material Erosion (1214); Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Hmrs parts replacement after 12 years.The wear of the components was confirmed during the surgery.The parts have all been replaced as requested.Some poly parts were severely damaged or disassembled into individual parts due to the difficult disassembly.Due to wear and tear, it was suspected that replacing the components would be prudent.
 
Additional Manufacturer Narrative
An event regarding wear involving a hmrs axle was reported.The event was confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that the device is worn along the circumference of the barrel.Damage appears consistent with contact against a hard object.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to wear of the components.Visual inspection of the returned device indicated that the device is worn along the circumference of the barrel.Damage appears consistent with contact against a hard object.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary/revision operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.It was reported that the patient was very active post-implantation.The patient's activity level may have contributed to the reported failure of the devices.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Hmrs parts replacement after 12 years.The wear of the components was confirmed during the surgery.The parts have all been replaced as requested.Some poly parts were severely damaged or disassembled into individual parts due to the difficult disassembly.Due to wear and tear, it was suspected that replacing the components would be prudent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HMRS AXIS PIN
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13767288
Report Number0002249697-2022-00395
Device Sequence Number5008706
Product Code KRO
UDI-Device Identifier07613327049930
UDI-Public07613327049930
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K122015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2013
Device Model Number6366-9-220
Device Catalogue Number63669220
Device Lot NumberCTD145A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/03/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexMale
-
-