Catalog Number JTS-DRIVE-UNIT
Device Problem Failure to Advance (2524)
Patient Problem Insufficient Information (4580)
Event Date 02/22/2022
Event Type  malfunction  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Failed lengthening reported.
 
Event Description
Failed lengthening reported.
 
Manufacturer Narrative
Please note that this event is related to a jts drive unit which is used in conjunction with the jts non-invasive extendible implant (k092138).Reported event: the reported device is a jts drive unit, sn: (b)(6).Method & results: device evaluation and results: visual inspection: visual inspection of the jts coil reported, showed overall general wear and tear on the product.Both the magnetic coil and the power unit console share the same serial number ¿(b)(6)¿ and both devices state the date of manufacture is ¿mar 2014¿.Dimensional inspection: not performed as not relevant to the reported event.Functional inspection: the drive unit mle3 lot: (b)(6) passed the annual maintenance/ function test.Material analysis: not performed as not relevant to the reported event.Clinician review: n/a.Device history review: review of the annual maintenance and functional inspection indicates the device conformed to requirements.Complaint history review: there have been no other events.Conclusion.The jts drive unit ((b)(6)) passed its annual maintenance test, confirming that drive unit is working accordingly.The exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, pathology reports, progress notes, x-rays and return of the device are needed to fully investigate the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JTS DRIVE UNIT SET
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK  WD6 3SJ
Manufacturer (Section G)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK   WD6 3SJ
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key13768196
MDR Text Key288908961
Report Number3004105610-2022-00026
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Reporter Country CodeBE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Followup
Report Date 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJTS-DRIVE-UNIT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-