| Model Number |
54840007550 |
| Medical Device Problem Codes |
Break (1069); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a patient during revision for po sterior lumbar fusion (plf).It was reported that the tulip head of the screw splayed during set screw implantation, the screw and set screw remains will be returned together.The tip of the screwdriver broke off during screw revision.No further complications or symptoms were reported.Additional information was received via manufacturer representative that the reported products were used in revision surgery.The reported products came in contact with the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: product# 54840007550 lot# h13a4788 after visual and optical examination and functional testing, it appears to be damage to the threads of the screw.There are some witness marks in the hex area.The screw and the broken hex driver tip where received separated and the driver tip was not received for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|