A nurse reported during the process of puncturing, the cannula was separated from the trocar and contaminated, hence could not be used normally while performing the vitrectomy procedure.The procedure was completed after changed to a new product.There was no patient harm.
|
A pak was received, however there were no trocar samples, at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.No trocar samples were received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Assemblies are 100% inspected for adhesive application during assembly.If adhesive application is incorrect the assembly is scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|