• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC Back to Search Results
Catalog Number 705.52P
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that during a vitrectomy surgery both ends of an ophthalmic micro scissor was stuck.The procedure was completed using an alternate product.No patient harm was reported.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Sample was not received by the investigation site for evaluation.The root cause cannot be identified conclusively because no picture or sample has been received.Investigation has been completed based on current information.Should sample returned, this file will be re-opened and failure analysis will be performed.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample is received with inner blister, outer blister and cover foil showing the lot information.The sample shows macroscopic signs of damage.One blade of the scissors was missing.The sample shows no signs of surgery residues.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was visually inspected.It was noticed that one blade of the scissors was missing.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.The customers complaint is confirmed the root cause cannot be identified conclusively because the sample has pass the 100% inspection.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.Additional information was provided in sections h.6 and h.10 a review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Sample was not received by the investigation site for evaluation.The root cause cannot be identified conclusively because no picture or sample has been received.Investigation has been completed based on current information.Should sample returned, this file will be re-opened and failure analysis will be performed.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP SCISSORS
Common Device Name
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13769192
Report Number3003398873-2022-00012
Device Sequence Number5032602
Product Code HNF
UDI-Device Identifier07612717070691
UDI-Public07612717070691
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number705.52P
Device Lot NumberF189274
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer05/17/2022
07/28/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date06/08/2022
08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-