• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH HMRS/G I-II. CIRCLIPS; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH HMRS/G I-II. CIRCLIPS; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6466-9-230
Medical Device Problem Codes Degraded (1153); Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Hmrs parts replacement after 12 years.The wear of the components was confirmed during the surgery.The parts have all been replaced as requested.Some poly parts were severely damaged or disassembled into individual parts due to the difficult disassembly.Due to wear and tear, it was suspected that replacing the components would be prudent.
 
Additional Manufacturer Narrative
An event regarding wear involving a hmrs circlips was reported.The event was not confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that no wear is present.There is nothing remarkable to note.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to wear of the components.Visual inspection of the returned device indicated that no wear is present.There is nothing remarkable to note.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary/revision operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.It was reported that the patient was very active post-implantation.The patient's activity level may have contributed to the reported failure of the devices.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Hmrs parts replacement after 12 years.The wear of the components was confirmed during the surgery.The parts have all been replaced as requested.Some poly parts were severely damaged or disassembled into individual parts due to the difficult disassembly.Due to wear and tear, it was suspected that replacing the components would be prudent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HMRS/G I-II. CIRCLIPS
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13769223
Report Number0002249697-2022-00398
Device Sequence Number4993463
Product Code KRO
UDI-Device Identifier04546540276520
UDI-Public04546540276520
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K122015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6466-9-230
Device Catalogue Number6466-9-230
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexMale
-
-