• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 SEPRAMESH IP; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 SEPRAMESH IP; SURGICAL MESH Back to Search Results
Catalog Number 5959480
Medical Device Problem Codes Defective Device (2588); Patient-Device Incompatibility (2682); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); Hernia (2240); Insufficient Information (4580)
Date of Event 01/11/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of unspecified bard/davol sepramesh ip and davol plug mesh (x2) on (b)(6) 2015 and/or (b)(6) 2018.As reported, the patient is making a claim for an adverse patient outcome against the sepramesh ip.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of unspecified bard/davol sepramesh ip and davol plug mesh (x2) on (b)(6) 2015 and/or (b)(6) 2018.As reported, the patient is making a claim for an adverse patient outcome against the sepramesh ip.Attorney alleges that the patient had subsequent surgical intervention due to the hernia mesh device.It is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2015 - patient was diagnosed with left inguinal hernia thereby underwent laparoscopic repair with implant of sepramesh ip (device 1).Per op notes, ¿an indirect hernia sac was noted and dissected free.A sepramesh ip (device 1) was placed and tacked using sutures.¿ (b)(6) 2018 - patient was diagnosed with recurrent left inguinal hernia with pain and right inguinal hernia thereby underwent open repair with implant of perfix plug (device 2 and 3).Per op notes, ¿an indirect and direct hernia sacs were noted on left and right inguinal regions.Both hernia sacs were reduced.Perfix plug (device 2) on left inguinal side and perfix plug (device 3) on right side were placed and sutured.¿.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges adverse patient outcome associated with the hernia mesh used to treat the patient.The patient's attorney alleges that the patient had subsequent surgical intervention; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-jul-2014) is considered to be a best estimate.Updated fields: a2, a4, b3, b4, b5, b7, d.4, d.6a (date of implant), g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEPRAMESH IP
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key13769293
Report Number1213643-2022-91986
Device Sequence Number3837084
Product Code FTL
UDI-Device Identifier00801741031816
UDI-Public(01)00801741031816(17)160728(10)HUYG1463
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K063739
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/28/2016
Device Catalogue Number5959480
Device Lot NumberHUYG1463
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer10/03/2025
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date10/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/15/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
Patient SexMale
Patient Weight73 KG
-
-