ZIMMER MANUFACTURING B.V. LINER 10 DEGREE ELEVATED RIM 32 MM I.D. FOR USE WITH 56 MM O.D. SHELL; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Event date: unknown day in (b)(6) 2019.Concomitant products: ref 2232-04-18 / lot 64013541, 63674177, 63853340, 63853474, 64013651 cerclage cables.Ref 6250-65-40 lot 63764148 bone screw.Ref 01.00102.614 lot 2810929 wagner sl femoral stem.The device will not be returned for analysis, as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2022 - 00057.
 
Event or Problem Description
It was reported the patient underwent initial right total hip arthroplasty.Subsequently patient had 3 dislocations and was revised approximately 10 years later.Patient has had 2 dislocations since revision with closed reductions approximately 6 months later.Components remain implanted.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the patient suffered from 3 (three) dislocations and was revised.Since then, the patient has experienced 2 (two) more dislocations that have been corrected with closed reductions.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined for the dislocations.It was identified that the head and stem used are not compatible together, however with the information provided it is unknown whether the off-label usage contributed to the dislocation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER 10 DEGREE ELEVATED RIM 32 MM I.D. FOR USE WITH 56 MM O.D. SHELL
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13770051
Report Number0002648920-2022-00056
Device Sequence Number4029219
Product Code LPH
UDI-Device Identifier00889024124769
UDI-Public(01)00889024124769(17)210531(10)63357496
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K990135
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model NumberN/A
Device Catalogue Number00631005632
Device Lot Number63357496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight109 KG
-
-