• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CRS RP INSRT SZ 5 8MM; KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CRS RP INSRT SZ 5 8MM; KNEE TIBIAL INSERT Back to Search Results
Model Number 1517-10-506
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Poly change from revision knee 3 years ago.No issue with poly, limited flexsion ~85° present in both knees (one dps / one exatech.Doi: unknown.Dor: (b)(6) 2022.Unknown knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d4 (catalog), d6b, g1 and h6 (impact code).Removed date of revision since there was no depuy product revised.
 
Event or Problem Description
Additional information was received indicated the following: 1.Was surgery time extended? if yes, what is the duration of the delay? no delay.2.What is the affected side involved in this event? left side.3.On 01-mar-2022 were any depuy products revised/explanted? no.4.How did the surgeon resolve the limited flexion issue? after initial exposure, scar tissue was extensively cleared and stretching of the remaining soft tissues resulted in an outcome rom and resting state the surgeon was happy with.5.Usage for the patient's revision procedure 3 years ago has been identified in the kit diary (see below).To your knowledge were both sides revised on (b)(6) 2019? no, only one side has a revision, the other side has more recently had a primary done using zimmer (mid-late 2021).I cannot speak to the double up with stems and sleeves.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CRS RP INSRT SZ 5 8MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13770319
Report Number1818910-2022-04720
Device Sequence Number3967137
Product Code NJL
UDI-Device Identifier10603295055914
UDI-Public10603295055914
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1517-10-506
Device Catalogue Number151710508
Device Lot Number8969262
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer03/18/2022
04/04/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date03/26/2022
04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL LT SZ 5 CEM; ATTUNE CRS RP INSRT SZ 5 8MM; ATTUNE DIST FEM AUG SZ 5 8MM; ATTUNE DIST FEM AUG SZ 5 8MM; ATUN FEM SLV M/L 30MM HALF POR; ATUN TIB SLV M/L 37MM FULL POR; ATUNE PRESSFIT STR STEM14X60MM; ATUNE PRESSFIT STR STEM16X60MM; ATUNE REV RP TIB BASE SZ 4 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-