• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SROM*STM ST,36+8L NK,20X15X165; S-ROM HIP SYSTEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SROM*STM ST,36+8L NK,20X15X165; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Model Number 52-3420
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Swelling/ Edema (4577)
Date of Event 10/24/2012
Type of Reportable Event Serious Injury
Event or Problem Description
Medical record received.After review of medical record, patient was revised to address painful left total hip, mom reaction.Revision notes stated that there was obvious distention of posterior rotators or capsule that was aspirated and found to be milky colored.There was obvious metal staining of the synovium.There was corrosion on the taper.Medical records indicate that the corrosion on the trunnion of the femoral stem was cleaned off.The stem and sleeve were well-fixed and therefore retained.Doi: (b)(6) 2007 - dor: (b)(6) 2012 (left hip).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS 1818910
700 orthopaedic dr
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13770532
Report Number1818910-2022-04732
Device Sequence Number17395369
Product Code LWJ
UDI-Device Identifier10603295171041
UDI-Public10603295171041
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2012
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2012
Device Model Number52-3420
Device Catalogue Number523420
Device Lot Number2334263
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2022
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/26/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINN CAN BONE SCREW 6.5MMX20MM; PINNACLE MTL INS NEUT36IDX54OD; PINNACLE SECTOR II CUP 54MM; S-ROM M HEAD 36MM +6; S-ROM*SLEEVE PRX ZTT, 20B-LRG
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexMale
-
-