MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; BLOOD GLUCOSE TEST STRIPS Back to Search Results
Model Number 7311
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family was the initial reporter , so personal information was not entered.No information was captured in as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that he attempted to perform blood glucose testing with the contour next link 2.4 meter and received error messages indicative of "used test strip".The customer then performed repeat testing and obtained lo reading, indicative of a reading below 20 mg/dl at 11:02 p.M.And 97 mg/dl at 11:03 p.M.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next link 2.4 meter and contour next test strips from lot #1epeg02b for evaluation.Therefore, an in-house testing was performed with the returned meter and test strips using blood sample, which gave a satisfactory performance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Common Device Name
BLOOD GLUCOSE TEST STRIPS
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION NUMBER: 1000134386)
2131-1 minamigata
toon ehime, 791-0 395
JA   791-0395
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key13770858
Report Number1810909-2022-00041
Device Sequence Number3822534
Product Code NBW
UDI-Device Identifier10301937311502
UDI-Public10301937311502
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2023
Device Model Number7311
Device Catalogue Number7311
Device Lot Number1EPEG02B
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer02/08/2023
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date02/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-