MAUDE Adverse Event Report: ZIMMER BIOMET, INC. NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 02/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after follow up on a total knee arthroplasty, it was discovered the patient had a tibia bone fracture and the tibial plate was subsiding.The patient was revised approximately two months post implantation; the femur, articular surface, and tibia were replaced.No additional information is available.
 
Additional Manufacturer Narrative
(b)(4).Associated product: femur item 42502806001 lot 64947129.Articular surface item 42512100710 lot 65178106.Foreign - (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No additional information was reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event or Problem Description
No additional information reported.
 
Additional Manufacturer Narrative
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Device history record was reviewed and no discrepancies relevant to the reported event were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified evidence of a periprosthetic fracture involving the medial tibial plateau extending to involve the medial aspect of the tibial component and the possibility of subsidence of the tibial component.A definitive root cause cannot be determined.Per package insert bone fracture is a known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13771422
Report Number0001822565-2022-00725
Device Sequence Number14704413
Product Code MBH
UDI-Device Identifier00889024510579
UDI-Public(01)00889024510579(17)310719(10)65016945
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42530007501
Device Lot Number65016945
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2022
Supplement Date Received by Manufacturer03/18/2022
04/07/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date03/24/2022
04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-