• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ VIO 27IN 0 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ VIO 27IN 0 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number VCP364H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: an image was received for review.Upon visual inspection of the image, five needles, one needle with attached suture, and some suture damage were noted.Also, the threads appear worn and with body fluids, the picture is not clear to determine the failure mode of the damaged suture.As part of our quality process, manufacturing records were reviewed for this lot's serial number and manufacturing standards were met prior to the release of this lot.No conclusion could be reached as to the cause of the reported complaint as the sample was not returned for testing.As part of quality process, all devices are manufactured, inspected and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2022-01828, 2210968-2022-01829 and 2210968-2022-01830.Trade name: irgacare®.Active ingredient(s): triclosan.Dosage form: suture/solid/parenteral.Strength: = 275 g/m.
 
Event or Problem Description
It was reported that a patient underwent a c-section procedure on (b)(6) 2022 and suture was used.During the procedure, 4 sutures got detached from the swage end of the needles during caesarean section.Another suture from the same lot broke during suturing.There were no patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 4/28/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.H3 investigational summary: two packets of product code vcp364h were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection to the samples no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand to detect any issue on the suture and no defects were observed during evaluation.A functional test was performed using an instron and the pull force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Trade name - irgacare®; active ingredient(s) ¿ triclosan; dosage form ¿ suture/solid/parenteral; strength ¿ = 275 g/m.Events reported via: 2210968-2022-01827, 2210968-2022-01828, 2210968-2022-01829 and 2210968-2022-01830.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 4/5/2022 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was received: the 5 impacted products were discarded by the surgeon.2 unopened representative samples from the same lot will be returned for analysis.Trade name - irgacare® active ingredient(s) ¿ triclosan dosage form ¿ suture/solid/parenteral strength ¿ = 275 ¿g/m.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 27IN 0 S/A CTX
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13771689
Report Number2210968-2022-01827
Device Sequence Number3967156
Product Code GAM
UDI-Device Identifier10705031052772
UDI-Public10705031052772
Combination Product (Y/N)Y
Initial Reporter CountryHK
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberVCP364H
Device Catalogue NumberVCP364H
Device Lot NumberPL5192
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer03/15/2022
04/05/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date04/05/2022
04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-