MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT231212
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 02/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The gore® excluder® aaa endoprosthesis instructions for use (ifu) states that adverse events that may occur and/or require intervention include, but are not limited to renal (e.G., artery occlusion).
 
Event or Problem Description
On (b)(6) 2022, the patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.It was reported that during the procedure, the trunk-ipsilateral leg endoprosthesis could not be reconstrained.The physician deployed the device, intentionally high, planning to reconstrain if needed.The device (the scallop at distal end of graft) partially covered the left renal artery.An angiograph was performed and showed flow into the left renal artery.The patient tolerated the procedure.The field sales associated stated ¿the clear dial became dislodged during shipping¿.
 
Additional Manufacturer Narrative
Added g3/g4, h1/h2 and h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
rachael chascsa
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13772519
Report Number3007284313-2022-01824
Device Sequence Number4994064
Product Code MIH
UDI-Device Identifier00733132622320
UDI-Public00733132622320
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/02/2024
Device Model NumberRLT231212
Device Catalogue NumberRLT231212
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/20/2022
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
Patient SexFemale
-
-