| Model Number |
B7016 |
| Medical Device Problem Code |
Unraveled Material (1664)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit is anticipated to return.A follow-up report will be provided upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: laser for stones.Complaint 1 of 2: (b)(4).Complaint 2 of 2: (b)(4).Event description: both these two axh sheath was used during a procedure.Outer piece spirrale.Kinked below.Additional information provided 15feb2022 via email from [name] non-conformance report: product kinked, broken, spiralled.Patient status: no patient injury.Type of intervention: ni.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of the sheath partially unraveling.Based on the condition of the returned unit, the reported event was most likely caused by contact with an instrument followed by a high force exerted on the unit.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.This event was initially reported based on the description of the event.However, based on the evaluation of the returned unit, applied medical determined that this event is not reportable as it is unlikely to cause or contribute to death or serious injury.
|
| |
|
Event or Problem Description
|
|
Procedure performed: laser for stones.Complaint 1 of 2:(b)(4).Complaint 2 of 2: (b)(4)- mfr 2027111-2022-00468.Event description: both these two axh sheath was used during a procedure.1.Outer piece spirrale.2 kinked below.Additional information provided 15feb2022 via email from [name] non-conformance report: product kinked, broken, spiralled.Additional information received from distributor via email on 10-mar-2022: no pictures were taken before the devices were sent to the manufacturer.Both access sheaths kinked during removal and one spiralled off during the procedure, as you can see in the returned product.Two other access sheaths were used and worked perfectly.Patient status: no patient injury.Type of intervention: two other access sheaths were used and worked perfectly.
|
| |
|
Search Alerts/Recalls
|