MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RMT231216
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 02/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The field sales associate reported the following: on (b)(6) 2014, the patient was implanted with a gore® excluder® aaa endoprostheses to treat an abdominal aortic aneurysm.The patient tolerated the procedure.It was reported that images (modality and date is unknown), revealed a possible type iii endoleak (tear in material) with aneurysm sac enlargement (amount of enlargement is unknown).On (b)(6) , 2022, the physician reintervened and implanted an aortic extender endoprosthesis and relined both limbs with contralateral leg endoprostheses.During the reintervention, no type iii endoleak was found, it is unknown what type of endoleak was present.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
Additional devices implanted and/or related to this event: pxc121200/ 12396311 udi (b)(4) and pxc121400/ 11538497 udi (b)(4).The gore® excluder® aaa endoprosthesis instructions for use (ifu) states that adverse events that may occur and/or require intervention include but are not limited to endoleak and aneurysm enlargement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY MOUNTAIN B/P
3750 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
rachael chascsa
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13773318
Report Number2017233-2022-02806
Device Sequence Number14700786
Product Code MIH
UDI-Device Identifier00733132610525
UDI-Public00733132610525
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model NumberRMT231216
Device Catalogue NumberRMT231216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/19/2022
Supplement Date Received by Manufacturer02/19/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/30/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
Patient SexMale
Patient Weight66 KG
-
-