| Model Number |
DFW150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); Halo (2227)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not returned for evaluation.Therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's right eye due to the patient was experiencing unbearable glare and halo 6 weeks post-operative (op and especially at night.The incision was enlarged.It was also reported that the patient was having difficulty driving at night.Pre-operative visual acuity is 20/60 dv cc (dv-distance vision, cc-with correction).The explanted iol was replaced with another j&j lens model zcu150 18.5 diopter (different model, same diopter).Post-operative visual acuity is 20/25 -1 dv and 2025 nv sc (nv-near vision, sc-without correction).Reportedly, there was no vitrectomy and no sutures were required.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional info: section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: mar 25, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: the suspect iol was received inside a specimen cup.The lens was removed and visual inspection revealed cosmetic issues and lens damage.No defects that could contribute to visual issues were observed.A product quality deficiency could not be determined.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
|
| |
|
Search Alerts/Recalls
|