MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20188
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the stent came off the balloon.The target lesion was located in the mildly tortuous and moderately calcified superior mesenteric artery.A 7.0mmx19mmx90cm express sd renal/biliary stent was advanced for treatment.However, the stent came off the balloon while trying to cross the lesion.The stent had to be snared and was pulled out through the sheath.The procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an express sd stent with a non-boston scientific sheath, 0.014" wire, and a snare.The balloon catheter did not return for analysis.Visual examination revealed that the stent is damaged.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no damage or irregularities.
 
Event or Problem Description
It was reported that the stent came off the balloon.The target lesion was located in the mildly tortuous and moderately calcified superior mesenteric artery.A 7.0mmx19mmx90cm express sd renal/biliary stent was advanced for treatment.However, the stent came off the balloon while trying to cross the lesion.The stent had to be snared and was pulled out through the sheath.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13776511
Report Number2134265-2022-02280
Device Sequence Number11739106
Product Code NIN
UDI-Device Identifier08714729390527
UDI-Public08714729390527
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K152607
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2024
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0027961075
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date03/15/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-