• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number 1076538
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.During device evaluation, particles inside flow path and all parts blower assembly, valve assembly, bottom enclosure, internal air cover, internal foam, blower box mount, right panel assembly, bottom label was observed.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported that required intervention to prevent permanent impairment/damage, occured in section b2.After review, it was determined that this options were incorrectly selected in section b2.Nothing such has occured to the patient.The report should have been sent for product problem only in section b1.In h6 type of investigation code was missing.The missing value is 10.In g4 section pma/510(k) number was incorrect the correct value is k121623.And b3, d9, e1, h3, d1, d2 are updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13777443
Report Number2518422-2022-10177
Device Sequence Number11213849
Product Code MNT
UDI-Device Identifier00606959060302
UDI-Public00606959060302
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/18/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1076538
Device Catalogue Number1076538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2015
Supplement Date Received by Manufacturer06/18/2015
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-