Corrected data: d9 additional data: h3, h6, h10 additional information was received from the reporter that the implant appeared to be broken before removal.Visual inspection found the nail to be fractured at the location of the screw holes on either side of the housing tube, where the screws are inserte d.Also, it was noted the broken ends were stretched, not straight.It is possible the nail was subjected to excessive bending, resulting in plastic deformation, which may have resulted in its breakage at the indicated position.The location of the failure (at the two holes on either side of the nail) is consistent with what would be expected from mechanical bending failure and fatigue.Based on these findings, it is possible that the fracture occurred due to stress generated from weight-bearing activities.Per the precice stryde instructions for use: "the precice stryde nail cannot withstand the stresses of full weight bearing.Patients should utilize external support and/or restrict activities as directed by the physician until consolidation occurs." production records were reviewed, and confirmed the device passed all inspections and acceptance tests prior to shipment.No material/manufacturing issue was found.
|