MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII C/R ART INS SZ7-8 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII C/R ART INS SZ7-8 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420514
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 05/28/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that after a bilateral tka on (b)(6) 2016, plaintiff presented anterior femoral resurfacing component lucency, arthritic varus knee deformity, increased of heterotopic ossification along the posterior knee and superior/inferior patella on the left side and pain in the groin.Plaintiff had a polyethylene insert fracture.Plaintiff underwent revision surgery on (b)(6) 2021 where the insert was explanted.The outcome of the plaintiffs is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation but the pictures were reviewed, and the broken device was confirmed.The clinical/medical investigation concluded that, the revision was due to mechanical instability/insert post component failure; however, the patient activity which led to the (b)(6) 2021 ed visit after a workout could not be ruled out as a potential contributing factor to the reported event.Additionally, the relatively stable lucencies and progression of heterotopic ossification and djd reported on x-rays may have also contributed to the reported pain, although not indicative of a component malperformance.The assessed patient impact was the reported symptoms, clinical findings, revision, and an anticipated transient post-op recovery.Further patient impact cannot be determined.No further medical assessment can be rendered at this time.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device revealed that fracture of implant has been identified in possible adverse effects section as a result of trauma, strenuous activity, improper alignment, or duration of service.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported event could be corroborated since the device broke, therefore, it could not function correctly anymore.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII C/R ART INS SZ7-8 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13777674
Report Number1020279-2022-01192
Device Sequence Number17077091
Product Code JWH
UDI-Device Identifier03596010208019
UDI-Public03596010208019
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420514
Device Catalogue Number71420514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Other;
Patient Age60 YR
Patient SexMale
Patient Weight109 KG
-
-