Model Number 66002880
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a black stain in the strip of leukostrip 13x102mm ctn 50.As this was noticed during set up, the treatment was performed, without any delay, using a s+n back-up device.Patient was not impacted.
 
Additional Manufacturer Narrative
Nternal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the product has been returned and evaluated, the event is confirmed.Visual inspection of the dressing has confirmed the reported contamination, this has been confirmed to be additional adhesive, this is not any other form of contamination.A documentation review has revealed previously cases with a similar nature, there are no open corrective actions.The manufacturing records confirm the device met the final product specifications and contain no contributory factors to the reported event, the associated risk files mitigate this event type.A review of the manufacturing process has revealed the probable cause.A probable cause includes the defect was not picked out from raw material prior to production, retraining has been conducted, to reduce the probability of future occurrences.No corrective actions are deemed necessary, at this stage, smith and nephew will continue to monitor for adverse trends relating to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEUKOSTRIP 13X102MM CTN 50
Common Device Name
TOPICAL APPROXIMATION SYSTEM
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13777974
Report Number8043484-2022-00102
Device Sequence Number16572735
Product Code MZR
UDI-Device Identifier05000223449186
UDI-Public05000223449186
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2023
Device Model Number66002880
Device Catalogue Number66002880
Device Lot Number2053A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-