| Model Number |
4054 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported an intermittent sensor.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported an intermittent sensor.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacuring narrative: other, other text: two sensors were returned and evaluated.During continuity testing, one sensor was found to have an open in cable on the emitter circuit and the other sensor had an open in cable on the detector circuit at the bend relief.During functional testing an error message was displayed on the lab monitor for both sensors.One sensor led did not illuminate; the other sensor led did illuminate.
|
| |
|