| Model Number |
4054 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported an intermittent sensor.No consequences or impact to patient were reported.
|
| |
|
Event or Problem Description
|
|
The customer reported an intermittent sensor.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the returned sensor was evaluated.An open in the cable on the emitter circuit at the bend relief was observed during continuity testing.This resulted in an error message to display on the lab monitor.The sensor led did not illuminate.
|
| |
|