Model Number Z9002
Medical Device Problem Codes Inaccurate Delivery (2339); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implant date: if implanted, give date: not applicable, as there is no indication that the lens was implanted.Explant plant: if explanted, give date: not applicable, as there is no indication that the lens was implanted.Device evaluated by mfr: the device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was partially inserted into patient's left eye, however, was removed as the lens was scratched.There was no patient injury and no other intervention was required.Another lens of same model and diopter was used as the replacement.No further information was provided.
 
Additional Manufacturer Narrative
Correction: section h6: medical device problem code: 2339 - inaccurate delivery code was incorrectly chosen in the initial report submitted (mfr (b)(4)).Therefore, it is being removed as it does not apply.Hence, this correction report is being submitted to capture this error.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: mar 23, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside of the original lens case.Visual inspection under magnification revealed that the lens was received crushed inside of the lens insert.The lens was cleaned, and scratches were observed on the lens.No further issues were identified with the lens.Conclusion: as a result of the investigation, the complaint issue was confirmed, however, it could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13779253
Report Number3012236936-2022-00639
Device Sequence Number19839291
Product Code HQL
UDI-Device Identifier05050474530737
UDI-Public(01)05050474530737(17)260804
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/04/2026
Device Model NumberZ9002
Device Catalogue NumberZ900200220
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2022
Supplement Date Received by Manufacturer04/01/2022
05/06/2022
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date04/26/2022
05/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-