Model Number 15
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code Angina (1710)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.The customer provided stryker with all the available patient information.Patient fields in which information is not provided were intentionally left blank.Stryker evaluated the customer¿s device and verified the reported issue.Stryker downloaded the electronic records of the device and observed the device experienced an inappropriate loss of power.Stryker isolated the cause of the reported issue to the power pcb assembly.Due to a part obsolescence, the device cannot be repaired.The customer was advised to replace their device.The device was sent to be scrapped.
 
Event or Problem Description
The customer contacted stryker to report that their device powered off by itself during a patient event.This issue is patient related; however there was no adverse patient outcome reported by the customer.
 
Additional Manufacturer Narrative
Stryker determined that the root cause of the reported issue was due to fretting corrosion on the battery wire harness contacts of connector j11 and the mating power pcb assembly contacts of connector p11.The fretting corrosion has been determined to cause an increase in contact resistance which can result in a sudden loss of device power under conditions of high current usage from functions such as printing a code-summary® record.The excessive voltage drop at the contacts triggers the power fail detector circuit on the power board, which causes the device to power cycle.
 
Event or Problem Description
The customer contacted stryker to report that their device powered off by itself during a patient event.This issue is patient related; however there was no adverse patient outcome reported by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key13779730
Report Number0003015876-2022-00371
Device Sequence Number1661328
Product Code MKJ
UDI-Device Identifier00883873912799
UDI-Public00883873912799
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer05/13/2022
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date05/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-