BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558340
Medical Device Problem Codes Burst Container or Vessel (1074); Leak/Splash (1354); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during a balloon dilatation procedure performed on (b)(6) 2022.During the procedure, the balloon leaked.The procedure was completed with another cre fixed wire rx dilation balloon.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.There were no patient complications reported as a result of this event.This event has been deemed reportable based on the investigation results of balloon burst.
 
Additional Manufacturer Narrative
Problem code: (b)(4).Investigation results: a visual and microscopic examination of the returned complaint device found that the balloon was ruptured, which confirms the reported event of balloon leaked.No damage found on the catheter of the device.The wire of the device was noted to be kinked.Based on the available information, the balloon was pre-inflated before inserting into the working channel that the condition can provoke the ruptured and the wire kinked found in the balloon, pre inflation of the balloon can provoke the balloon to loose its original form and cause friction inserting it on the working channel.Therefore, the most probable root cause is failure to follow instruction.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was not used per the directions for use (dfu)/product label as the balloon was pre-inflated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13780771
Report Number3005099803-2022-01225
Device Sequence Number13849175
Product Code KNQ
UDI-Device Identifier08714729201991
UDI-Public08714729201991
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K971320
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2023
Device Model NumberM00558340
Device Catalogue Number5834
Device Lot Number0026961705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2022
Initial Report FDA Received Date03/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-