| Model Number |
Z9002 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) was inserted into patient's left eye, however, was removed during the same procedure as the lens tore upon insertion.Another lens of same model and diopter was used as the replacement.There was no patient injury and no intervention was required.The patient was doing good post operation.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Implant date: if implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.Explant date: if explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.Device evaluated by mfr: the device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes section d9: returned to manufacturer on: mar 25, 2022 section h3: device evaluated by manufacturer: yes device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the lens was cut in half.The lens was cleaned and no further issues were observed.Due to the returned condition of the lens (cut in half) no further product evaluation could be performed.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed one additional investigation request (ir) for this po has been received, however, complaint was not confirmed based on the investigation and no escalation was required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|