BOSTON SCIENTIFIC CORPORATION CRE PULMONARY; BRONCHOSCOPE ACCESSORY Back to Search Results
Model Number M00550310
Medical Device Problem Codes Material Puncture/Hole (1504); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device code: (b)(4).The complainant indicated that the device has been disposed and is not available for return; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was used in the lung during an airway balloon dilation procedure performed on (b)(6).2022.During the preparation, it was noticed that the tip of the balloon had a pinhole leak.The procedure was completed with another cre pulmonary dilatation balloon.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE PULMONARY
Common Device Name
BRONCHOSCOPE ACCESSORY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13783799
Report Number3005099803-2022-01309
Device Sequence Number11191254
Product Code KTI
UDI-Device Identifier08714729456193
UDI-Public08714729456193
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K023337
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2024
Device Model NumberM00550310
Device Catalogue Number5031
Device Lot Number0028495004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2022
Initial Report FDA Received Date03/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-