H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one bard magnum biopsy needle for evaluation.During visual evaluation, the device appeared clean and the spacer attached to the needle at the hubs.Distal tip of stylet appeared bend.It was also noted that the sample notch was bent forwards when positioned on a flat surface.The functional testing was not performed, due to the condition of sample.Upon, dimensional observation, the notch bend was measured and found to be not within specification.Then the notch thickness was measured and found to be within specification.Therefore, the investigation is confirmed for the reported needle bent as the distal tip of stylet appeared bend.However, the investigation is inconclusive for the reported difficult to remove as the functional testing was not performed.Three electronic photos were provided and reviewed.All the three photos shows only the half part of the magnum needle and it was noted that the distal tip of stylet appeared bent.Based on the photo review, the reported bent needle can be confirmed.Therefore the investigation is confirmed for the reported bent needle and inconclusive for the reported difficult to remove.A definitive root cause for the alleged bent needle and difficult to remove could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 07/2022), g3, h6 (method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|