Additional Manufacturer Narrative
|
|
Image review: one photo was provided by the account for review.The device is in a clear biohazard bag.A detachment is visible to the balloon and shaft of the catheter.The balloon material is deformed.The material at the detachment site of the shaft appears to be discolored and stretched.Stretching is also evident to the shaft immediately proximal to the tip.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use the in.Pact acv access dcb during procedure to treat av fistula.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.It was reported that balloon was broken.It is unknown if the device or component detached, cracked, or fractured.The detached portion of device did not remain in patient.The procedure was completed successfully.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: stenosed av fistula, no calcium.There were no issues noted when removing the device from the hoop/tray.The device was prepped per the ifu.The balloon was inflated according to ifu.No resistance was encountered when advancing the device.The device did not pass through a previously-deployed stent.It is reported balloon burst during inflation.Balloon burst in av fistula (brachial artery) detached in sheath.When it was being retrieved, it broke off in the sheath and was retrieved without issue.Device was removed manually through sheath.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|