MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT EZ; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT EZ; BLOOD GLUCOSE METER Back to Search Results
Model Number 7252
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that her blood glucose readings were not being stored in the memory of the contour next ez meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
 
Additional Manufacturer Narrative
A repeat in-house testing was performed due to the error in laboratory testing.The returned contour next ez meter was tested with the returned contour next test strips using blood sample, which gave a satisfactory performance.The blood glucose readings obtained during the in-house testing were properly stored in the meter's memory.The device manufacture date was inadvertently changed in the 2nd supplemental.Section h4 has been updated with the correct device manufacturing date which matches with the initial report.
 
Additional Manufacturer Narrative
A repeat in-house testing was performed due to the error in laboratory testing.The returned contour next ez meter was tested with the returned contour next test strips using blood sample, which gave a satisfactory performance.The blood glucose readings obtained during the in-house testing were properly stored in the meter's memory.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next ez meter and contour next test strips for evaluation.An in-house testing was performed with the returned meter and test strips using blood sample, which gave a satisfactory performance.The blood glucose readings obtained during the in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT EZ
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION #: 3004167884)
110 oaza-inoshiri-aza
-nishiueno, wakimachi
mima tokusima, 779-3 603
JA   779-3603
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key13789867
Report Number1810909-2022-00049
Device Sequence Number3905710
Product Code NBW
UDI-Device Identifier00301937252013
UDI-Public00301937252013
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111268
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number7252
Device Catalogue Number7252
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/03/2022
Supplement Date Received by Manufacturer02/08/2023
02/23/2023
03/17/2023
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date02/21/2023
03/06/2023
03/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-