Model Number 6260-5-132
Medical Device Problem Codes Corroded (1131); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 01/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her right hip on or about (b)(6) 2017 and was revised on (b)(6) 2021.It is further alleged that she suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level, wear (trunnionosis), altr involving a metal head was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-medical records received and evaluation: a review of the provided medical information by a clinical consultant indicated: "review of the records confirms the index surgery was completed with a cocr v40 lfit femoral head.The surgeon in his revision operative note states that the patient had a pseudotumor and altr and believed there was evidence of trunionosis which he felt to be the cause.He performed a head revision only using a ceramic head and metal sleeve.It is confirmed that a revision tha was performed for what was thought to be trunnionosis with an altr and pseudotumor.Other than the description in the operative record no further evidence of elevated serum ion levels, altr and pseudotumor was presented.The root cause for the trunnionosis leading to these issues and the revision surgery can not be determined from the limited information available." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in her blood.A review of the provided medical information by a clinical consultant indicated: "review of the records confirms the index surgery was completed with a cocr v40 lfit femoral head.The surgeon in his revision operative note states that the patient had a pseudotumor and altr and believed there was evidence of trunionosis which he felt to be the cause.He performed a head revision only using a ceramic head and metal sleeve.It is confirmed that a revision tha was performed for what was thought to be trunnionosis with an altr and pseudotumor.Other than the description in the operative record no further evidence of elevated serum ion levels, altr and pseudotumor was presented.The root cause for the trunnionosis leading to these issues and the revision surgery can not be determined from the limited information available."the exact cause of the event could not be determined because insufficient information was provided.The medical records provided for review were insufficient to confirm the reported events or determine root cause.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her right hip on or about (b)(6) 2017 and was revised on (b)(6) 2021.It is further alleged that she suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM STD V40 TAPER VIT HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NJ NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13790627
Report Number0002249697-2022-00403
Device Sequence Number19839391
Product Code JDI
UDI-Device Identifier07613327012873
UDI-Public07613327012873
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Model Number6260-5-132
Device Catalogue Number6260-5-132
Device Lot Number51963303
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date03/16/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/02/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-