| Model Number |
9735542 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding a thermal therapy system being used in a soft tissue ablation (neuro) procedure.It was reported that the catheter charred.The manufacturer representative (rep) stated that after a case when they were removing the lasers where 3 were placed, it was noticed that one of the lasers were charred.There rep believes this was the 3rd laser that was ablated.There was no reported delay to the case.No reported impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: other relevant device(s) are: product id: 9735560, serial/lot #: unknown, product id: m450001b022, version #: (b)(4).The fiber was returned to the manufacture for evaluation.After visual/physical examination the reported issue was confirmed.The tip of the returned catheter was burnt.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the software investigation found that the reported event was unrelated to a software issue.The software functioned as designed.H6: fdm b01, fdr c19, and fdc d14 are applicable to the software analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|