| Model Number |
RONYX35015UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
One resolute onyx coronary drug eluting stent was attempted to be used to treat a severely tortuous and calcified lesion in the ostium of the circumflex (cx).A non-medtronic guide wire was used.The lesion was pre-dilated a few times using a 3.0x15mm euhpora balloon.Ivus was also used.The stent failed to cross the lesion.More euphora balloons were used.It was reported that on the third attempt to pass the stent, stent dislodgement occurred in the left main (lm) and cx.The physician passed a 2.0x12mm euphora balloon through the dislodged stent and inflated it.A non-medtronic stent was attempted to be used to cover the lesion but failed to cross.Another resolute onyx stent was able to cross the dislodged stent and was deployed.A 4.0x12mm nc euphora balloon was used to dilate both stents.No further patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the device was inspected before use with no issues noted.The was device prepped per ifu with no issues noted.The device did not pass through a previously deployed stent or cell of a stent.Resistance was not noted advancing the device.Excessive force was not used.Resistance was not noted during withdrawal of the device and excessive force was not used.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the device was received for analysis.Kinks were evident on the hypotube.The stent was not present on the balloon and did not return for analysis.Crimp impressions were visible on the exposed balloon surface.The balloon folds remained intact.The inner lumen patency was verified with a 0.015 inch mandrel.Deformation evident to the distal tip.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|