Model Number 66800040
Medical Device Problem Codes Disconnection (1171); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that, during debridement, the connecting tube of the versajet exact assy fell off.Procedure was resumed, after a +31 minutes delay, using a back up device.No other complications reported.
 
Additional Manufacturer Narrative
The device was returned for evaluation, with no faults found the event cannot be confirmed.Visual inspection noted no defects, the returned device had the saline supply tube securely attached to the orange key.A functional test was conducted, the feed line was pulled, with moderate pressure, to attempt to disconnect it from it's pump cartridge fitting, the feed line did not disconnect with moderate pressure.Greater pressure was applied and the feed line dislodged from the pump cartridge.The feed line appears to have been correctly attached to the feed line fitting within the pump cartridge during manufacture.A review of the supplied image shows a handpiece with the saline hose detached.Although the sample received contradicts this event.A documentation review has confirmed this device met the final product specifications.The review has revealed previous instances of this nature, no manufacturing quality concerns have been observed and no open corrective actions.A probable root cause may include that great pressure was placed upon the saline tube causing the tube to become dislodge from the orange key.No manufacturing quality concerns have been observed, therefore no corrective actions are deemed necessary, smith and nephew will continue to monitor for adverse trends relating to the reported allegation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET EXACT ASSY, 15 DEGREE X 14MM
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13791337
Report Number8043484-2022-00103
Device Sequence Number1688809
Product Code FQH
UDI-Device Identifier00040565124674
UDI-Public00040565124674
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K143115
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2022
Device Model Number66800040
Device Catalogue Number66800040
Device Lot Number50877879
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-