| Model Number |
MC2-5590S |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient's birth year is (b)(6).The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that the rod could not be distracted anymore due to the end cap loosening of the actuator.The rod was explanted and a fusion took place from t2-l4.No patient adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint 1 of 2.Visual inspection of the returned device revealed that the rod had been partially distracted.X-ray images of the internal components showed no damage, and confirm the rod was partially distracted.Additionally, x-ray images of the internal components revealed that the threaded cap was seated properly between the distraction rod and the housing tube.No issue was observed with the end cap.The rod was functionally tested; during testing the rod was able to be distracted and retracted with the external remote controller (erc) as well as with the manual distractor.The length as received was measured at 310.54 mm.The minimum and maximum lengths were measured at 309.80 mm and 357.10 mm respectively; giving a total stroke measurement of 47.30 mm, which meet the specification.Distraction force testing was measured at 45.20lbs, which meets specification.Reported failure event cannot be duplicated or confirmed.A review of the dhr documents indicated that the rod has been manufactured by the specified requirements at that time and met all of the required quality inspections before shipment.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|