MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5590S
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient's birth year is (b)(6).The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod could not be distracted anymore due to the end cap loosening of the actuator.The rod was explanted and a fusion took place from t2-l4.No patient adverse event was reported.
 
Additional Manufacturer Narrative
Complaint 1 of 2.Visual inspection of the returned device revealed that the rod had been partially distracted.X-ray images of the internal components showed no damage, and confirm the rod was partially distracted.Additionally, x-ray images of the internal components revealed that the threaded cap was seated properly between the distraction rod and the housing tube.No issue was observed with the end cap.The rod was functionally tested; during testing the rod was able to be distracted and retracted with the external remote controller (erc) as well as with the manual distractor.The length as received was measured at 310.54 mm.The minimum and maximum lengths were measured at 309.80 mm and 357.10 mm respectively; giving a total stroke measurement of 47.30 mm, which meet the specification.Distraction force testing was measured at 45.20lbs, which meets specification.Reported failure event cannot be duplicated or confirmed.A review of the dhr documents indicated that the rod has been manufactured by the specified requirements at that time and met all of the required quality inspections before shipment.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise #100
aliso viejo, CA 92656
8583448112
MDR Report Key13791432
Report Number3006179046-2022-00156
Device Sequence Number11213954
Product Code PGN
UDI-Device Identifier00812258022662
UDI-Public812258022662
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5590S
Device Lot Number9030703
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-