• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR CERTAS PLUS INLINE WITH SIPHON; CERTAS PLUS W/ SG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR CERTAS PLUS INLINE WITH SIPHON; CERTAS PLUS W/ SG Back to Search Results
Model Number 82-8804PL
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Cerebral Ventriculomegaly (2133)
Date of Event 03/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A physician reported a certas plus valve was implanted on (b)(6) 2021 with setting 4.The patient presented with increase in ventricle size.The physician tried to reprogram the valve to 3 with a etk and the valve stayed at 4.After many attempts including with the basic tool kit and low profile tool, the setting remained at 4.The large magnet couldn¿t move the setting off of 4.X-ray confirmed setting is stuck at 4.The valve was explanted and replaced on (b)(6), 2022.
 
Additional Manufacturer Narrative
The certas valve (id 828804pl) was returned for evaluation.Failure analysis - the position of the cam when valve was received was at setting 4.The valve was visually inspected; needle hole was noted in the needle chamber.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer could be due to biological debris and protein build up interfering with the valve mechanism, but at the time of investigation no programing issues were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAS PLUS INLINE WITH SIPHON
Common Device Name
CERTAS PLUS W/ SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13791619
Report Number3013886523-2022-00093
Device Sequence Number13904693
Product Code JXG
UDI-Device Identifier10886704079040
UDI-Public10886704079040
Combination Product (Y/N)N
PMA/510(K) Number
K143111
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number82-8804PL
Device Catalogue Number828804PL
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-