| Model Number |
40000006 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the patient was having an icd lead revision and during the procedure the patient required cardio version x2 and defibrillation x2.After the first defibrillation it was found that the pacing pad lead rhythm was absent on the phillips monitor.The pacing lead attached to the pad on the patient's back had dislodged from the pad.New pads were quickly applied and the second defibrillation was effective with new pads.The patient was not harmed but it was a close call.The patient responded well to defibrillation and there was no unexpected or prolonged care.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction made to h6 health effect - impact code to reflect "no health consequences or impact".
.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the issue reported.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.There were no deviations or nonconformances initiated with this lot code.A sample was not received for the investigation.Because a sample was not returned, no physical product examination was possible to evaluate the wire connection and properties of the electrode; therefore, the reported condition could not be confirmed at this time.Without a sample that displays the reported condition or a sample in which the reported condition can be duplicated, the root cause cannot be positively attributed to a product design or manufacturing problem.Current manufacturing practices require production to be sampled and inspected through a series of functional and visual testing based on valid acceptable quality limit sampling plans to ensure connectivity and shock delivery.The testing performed for this product lot was within specification.Since this complaint is unconfirmed, no complaint trend exists, and a specific root cause for the occurrence cannot be identified, no corrective action is required at this time.If a sample is received at a later date, the complaint will be reopened, and the investigation will be updated to reflect our findings.This complaint will be recorded for tracking and trending purposes.
|
| |
|
Search Alerts/Recalls
|