• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CRS FEMORAL RT SZ 7 CEM; KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CRS FEMORAL RT SZ 7 CEM; KNEE FEMORAL Back to Search Results
Model Number 1504-40-207
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision completed due to pain and loosening.It is unknown which products and interfaces were loose.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received: 1.Were all the reported products revised? if no, please specify which ones were revised.Yes.2.Was there a surgical delay? if yes, what was the duration of the delay? no.3.It was indicated that there was loosening.What component was loose? on what interface did the loosening occur? tibia.4.Was the bone cement/s used depuy manufactured? if yes, please provide the quantity, product and lot numbers.No.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.A photo was provided for review.Review of the photo found no evidence of device failure or malfunction, which could contribute to reported event.Depuy synthes considers, the investigation closed at this time.Should additional information be received, the information will be reviewed.And the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CRS FEMORAL RT SZ 7 CEM
Common Device Name
KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13791976
Report Number1818910-2022-04814
Device Sequence Number1805761
Product Code JWH
UDI-Device Identifier10603295042303
UDI-Public10603295042303
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K160700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1504-40-207
Device Catalogue Number150440207
Device Lot NumberHK4748
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2022
Supplement Date Received by Manufacturer05/08/2022
05/26/2022
Initial Report FDA Received Date03/17/2022
Supplement Report FDA Received Date05/18/2022
05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE POS FEM AUG SZ 7 8MM; ATTUNE PS RP INSRT SZ 7 7MM; ATUN FEM SLV M/L 45MM HALF POR; ATUN TIB SLV M/L 45MM FULL POR; ATUNE PRESSFIT STR STEM12X60MM; ATUNE PRESSFIT STR STEM14X60MM; ATUNE REV RP TIB BASE SZ 5 CEM; CEMENT MFG AND PRODUCT NAME UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
Patient Weight81 KG
-
-