| Model Number |
MLCT4S20 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Our investigation is ongoing.A follow up report will be submitted when the investigation is complete.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
During midline insertion on right brachial vein, guidewire well positioned with ultrasound, after peel-away insertion, guidewire removal was limited by obstacles.All peel-away plus guidewire was removed and the tip of guidewire was coiled.After small bleeding stop, new set was used and placed correctly.
|
| |
|
Additional Manufacturer Narrative
|
|
The device involved in the incident was not returned for evaluation and no photographs were provided.A record review request was sent to the device contract manufacturer.The inspection method performed was a device history record (dhr) review.The review revealed the following: - no internal nonconformance reports were found on the batch reported.- personnel was properly trained.- the reviews of the dhr demonstrate compliance and conformance to applicable procedures and specifications.- pull test results were verified and it was confirmed the batch meet the tensile specifications.- there is no evidence to confirm that this event is caused by manufacturing deviations.- therefore, it is considered a non-manufacturer-related issue.- the root cause is indeterminable.A possible cause of the knotted tip could be a back-and-forth motion during insertion.This can cause the wire to fold back on itself and result in a knotted tip.This may also cause the wire to be difficult to remove, resulting in applying more forces than the device is designed to withstand in order to remove it.
|
| |
|
Search Alerts/Recalls
|