MICROVENTION, INC. AZUR CX 18D-LONGER; VASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 45-780413-L
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not available for return to the user facility.It was not returned to the manufacturer for analysis.The alleged product issue could not be confirmed.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Event or Problem Description
It was reported that an embolization coil implant was noticed to be detached from the delivery pusher within the microcatheter.The device was removed with the microcatheter in its entirety.There was no reported intervention of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR CX 18D-LONGER
Common Device Name
VASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key13792192
Report Number2032493-2022-00088
Device Sequence Number1805195
Product Code KRD
UDI-Device Identifier00810170018299
UDI-Public(01)00810170018299(11)200430(17)260228(10)2103095ZA
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K123384
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model Number45-780413-L
Device Catalogue Number45-780413
Device Lot Number2103095ZA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2022
Initial Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1