Product complaint number: (b)(4.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, and the following was obtained: what tissue was being approximated or sutured when it broke? muscle tissue what was used to complete the procedure? 2 polysorb from covidien.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent related to: 2210968-2022-01862.
|
|
Product complaint # (b)(4).Date sent to the fda: 4/4/2022.H6 component code: g07002 no device problem found.H3 investigational narrative: the product was returned to ethicon inc for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that an empty opened folder and two suture pieces of product code j649g were received for evaluation.The sutures pieces present body fluids and the ends aren't broken, their cut appears to be by de use of a surgical instrument.Due to the condition of the sample the functional test can¿t be performed.In addition, a suture piece present a knot formed with the surgical technique.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.No conclusion could be reached as to what caused the reported complaint.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.The manufacturing records couldn't be reviewed as the batch number is unknown.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational narrative: one packet of product code j649g was returned to ethicon inc for analysis.Upon initial inspection of the sample, no external damages were observed on the packet.In order to evaluate the conditions of the returned sample, the packet was opened, and no defects were detected.The swage and attachment area were noted to be as expected.The suture wad dispensed without problems and examined along the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.Functional test was performed, and the tensile strength result was above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|