| Model Number |
H7493941816220 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Synergy xd mr us 2.25 x 16 mm stent delivery system (sds) was returned for analysis.Examination of the crimped stent via scope found evidence of damage with stent struts from the proximal stent region lifted and pulled distally.The undamaged crimped stent outer diameter was measured by snap gauge and the result was within max crimped stent profile measurement.The balloon cones were reviewed via scope, and no issues were noted.The balloon wings were tightly wrapped and evenly folded.The balloon had not been subjected to positive pressure.Examination of the tip via scope found no issues.A visual and tactile examination of the hypotube found no issues.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and a tactile examination was performed.No issues were noted.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 09-mar-2022.It was reported that crossing difficulties were encountered.The target lesion was located in a tortuous and complex vessel.A 2.25 x 16mm synergy xd drug eluting stent was advanced for treatment but failed to cross the lesion.The procedure was completed with another of same device.No patient complications were reported.However, returned device analysis revealed stent damage.
|
| |
|