| Model Number |
FT15HN60NSF940N |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the cuff leaked after four hours.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could not be duplicated.The device passed all functional tests.Dhr review was done, no issues related to the original complaint were found.
|
| |
|