| Catalog Number |
2H8519 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter phone no.(b)(6).Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a clearlink system non-dehp continu-flo solution set leaked from "the tube".The event occurred during prime.Therefore, there was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.Visual inspection was performed using the naked eye which revealed a cut in the tubing.Functional testing was performed including clear passage and pressure testing which revealed a leak from the affected area.The reported condition was verified.The cause of the condition was due to manufacturing; potentially related to the coiling or packaging step.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|